Model:ROC-5200 Reverse osmosis controller
Typical applications
Laboratory highly purified water preparation system integrated all-in-one PC reverse osmosis water program controller and water conductivity/resistivity monitoring
Technical data
Reverse osmosis collect: Raw water without water protection、High pressure water full protection、 Running reset
Reverse osmosis control points: high-pressure pump、circulating pump、flush valve、RO outlet valve、Disinfection valve、Disinfection discharge valve、DI(pure water) outlet valve、Inner circulation valve
Measurement to collect: Raw water conductivity acquisition、Level 1 water production conductivity acquisition、Secondary water production conductivity acquisition、Temperature gathering
Test project: conductivity/ resistivity/ temperature
Electrode choosing: Measuring 1.0 cm - 1 with electrical conductivity measuring resistivity electrode with 0.05-1 cm
Measurement range: Raw water conductivity 0~2000ppm Match 1.0 cm - 1 electrode
RO Water production electric conductivity 0~200ppm Match 1.0 cm - 1 electrode
DI Water production resistivity 0~18.2MΩ·cm(Match 0.05-1 cm electrode)
Accuracy: ≤ 1.5 %( FS)
Temperature compensation: In 25 ℃ as a benchmark, conductivity automatic digital compensation, resistivity automatic simulation compensation
Working conditions:ambient temperature:0~50℃;relative humidity:≤85%RH
Controlled contact KSK: contactless power relay
Data update rate: ≤1 second/times
Conductance cell working pressure: 0~0.5Mpa
Dimension: 160mm×100mm×40mm(H×W×D)
SlService voltage: DC 24V±4V
Slot dimension: 57mm×106mm (H×W)
Installation: Machine fixation
|