Model :CM-230/330 Conductivity monitors/ Measurement instrument
Features and Applications
1、Auv industrial online conductivity surveillance, Measurement instrument, small volume and low price
2、Adopt the artificially hardware Settings of switch range and Conductance pool constant input
3Simulation of temperature compensation
4、Current output without isolation
5、Extensive use of monitor with Pure water manufacturing, Cooling towers and General water monitoring.
Technical Dat
|
CM-230 |
CM-330 |
Measurementrange |
0~20、0~200、0~2000μS;
0~20mS/cm |
0~20、0~200、0~2000μS;
0~20、0~200、0~2000ppm;
0~20mS/cm |
Accuracy |
1.5%(FS) |
Temperature compensation |
In 25 ℃ as a benchmark, simulation of temperature compensation |
medium temperature |
5~50℃ |
Electrode coefficient |
1.0cm-1 |
Display Modes: 显示型态;显示模式 |
3 balance point a backlighting LCD |
Current output |
isolation 4 ~ 20mA |
— — — |
Control output |
— — — |
ON/OFF Upper limit relays |
Power supply |
AC 220V±10% 50/60Hz |
Working conditions |
ambient temperature:0~50℃ relative humidity:≤85% |
Dimension |
48×96×100mm(H×W×L) |
Slot dimension |
45×92mm(H×W) |
Installation |
panel mounted
Purchase Units
Basic parts:
CCT-8551 Intelligent Network Conductivity/ Resistivity Transmitter Controller(Single channels)
Sensors selection:
CON5121-13(0.01cm-1) two polar plate type titanium sensor used for high pure water measure
CON3123-13(0.1cm-1) two polar plate type st.st.316L sensor used for RO industry pure water measure
CON2124-13(1.0cm-1) two polar plate type graphite sensor used for industry pure water and municipal drinking water measure
CON2126-13(10.0cm-1) two polar plate type graphite sensor used for waste water and high salinity water measure
CON2227-13 Four sensors used for wide range variety media |
|